Terminated

XoranConnect, software version 4.1.0.6, Xoran Technologies, Ann Arbor, MI. A web-based software which provides online multi-planar image viewing, offsite data storage, administrative reporting tools, and teleradiology workflow. Server designed to store and serve stored imaging content on a RAID5 storage device.

FDA Recall Z-1577-2009 · Posted 2009-07-09

Recall StatusTerminated
Product CodeLLZ
Date Initiated2009-02-03
Date Terminated2009-11-19
Product Quantity14

Reason for Recall

An incorrect patient name label may appear on the screen displaying the selected patient's scan.

Firm Action

The firm released software version 4.1.0.7 by uploading it through a remote software upgrade on February 3, 2009. User accounts were notified by letter dated February 9, 2009. Direct questions to Xoran Technologies at 1-800-709-6726.

Root Cause

Software design

Distribution Pattern

Nationwide Distribution (Alabama, Arizona, Colorado, Georgia, Illinois, Louisiana, New York, North Carolina, Oklahoma, Rhode Island, Tennessee, Texas and Utah).