Zenition 70. Product Code 718133. The devices are used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the devices.
FDA Recall Z-1581-2025 · Posted 2025-04-21
Advertisement
Recalling FirmPhilips North America Llc
Recall StatusOpen, Classified
Product CodeOWB
Date Initiated2025-02-25
Reason for Recall
A wireless foot switch pedal may get stuck in the active position when the user releases the pedal,
resulting in the emission of unintended radiation.
Firm Action
Philips will send a customer notification to the existing users of the affected systems to alert them
of the issue and provide the actions that should be taken in order to prevent risks to patients. Philips will contact customers to schedule a visit to inspect the foot switch and provide an
addendum to the Instructions for Use (IFU) of the system detailing the appropriate cleaning and
maintenance instructions.
Philips proposes sending the Electronic Product Radiation Defect notification to customers
immediately upon FDA approval of the corrective action plan, and completion of the remaining
activities within 4 months of FDA approval of the corrective action plan.
This corrective action does not require any customer transportation reimbursement.
The following statement is included in the customer communication:
The manufacturer will, without charge, remedy the defect or bring the product into compliance
with each applicable Federal standard in accordance with a plan to be approved by the Secretary
of Health and Human Services.
Draft of letter provided. Philips commits to notifying the FDA on a monthly basis providing progress with this field
correction.
Root Cause
Labeling design
Distribution Pattern
US
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.