Cardiomedics CardiAssist External Counterpulsation System, circulatory assist device, Series 4000.
FDA Recall Z-1582-2009 · Posted 2009-07-07
Advertisement
Recalling FirmCardiomedics Incorporated
Recall StatusTerminated
Product CodeDRN
Date Initiated2009-04-09
Date Terminated2012-08-27
Product Quantity1
Reason for Recall
Cuff may underinflate, or device may stop working during use. Cardioassist software version 4.3.1 utilized a signal from the plethysmograph that was found not to be sensitive enough due to the shape of the waveform in certain circumstances where there was excessive motion artifact.
Firm Action
Service visit to customer in May 2009 to upgrade and validate CardioAssist 4000 software to version 4.3.5.
Root Cause
Software design
Distribution Pattern
Florida