DeRoyal(R) SURGI MATE(R) Electrosurgical Pencil, Button Switch, with 15 ft (4.6 m) Cord and Holster, REF 88-001502 Product Usage: general surgical

FDA Recall Z-1583-2018 · Posted None

Recall StatusTerminated
Product CodeGEI
Date Initiated2018-04-23
Date Terminated2018-11-09
Product Quantity1800 pencils

Reason for Recall

Non-sterile product was packaged inside a case that was labeled as sterile.

Firm Action

DeRoyal sent an Urgent Recall Notice letter dated April 23, 2018 to affected customers. The letter identified the affected products, problems and actions to be taken. The recall notices requested destruction of the product at the consignee level. Distributors were asked to notify their consignees or to provide the list to DeRoyal for notification. For questions contact your DeRoyal Representative, customer service at 888-938-7828 or by email at [email protected]

Root Cause

Process control

Distribution Pattern

Honduras, Belgium, and The Netherlands