Open, Classified

Zenition 70. The device is to be used in health care facilities both inside and outside the operating room, sterile as well as non-sterile environment in a variety of procedures.

FDA Recall Z-1583-2025 · Posted 2025-04-21

Recall StatusOpen, Classified
Product CodeOWB
Date Initiated2023-07-20
Product Quantity12 systems

Reason for Recall

Philips has become aware of the potential for unintended radiation exposure with Zenition 70 systems with Vascular Extension option and a Wireless Foot Switch.

Firm Action

Philips will send a customer notification to the existing users of the affected systems to alert them of the issue and provide the following actions that should be taken to prevent risks to patients: Keep the Electronic Product Radiation Correction letter with the documentation of the system. Where possible, place the Wireless Foot Switch within a 3 meter range from the C-arm stand to minimize as much as possible the potential delay of 72 milliseconds. Circulate this Electronic Product Radiation Correction letter to all users of the system so that they are aware of the issue. Philips will contact customers to schedule a visit to arrange for a Field Service Engineer visit to update the Software free of charge. (Philips reference 2024-IGT-PUN-007).

Root Cause

Software design

Distribution Pattern

US Nationwide distribution