Bard Luminexx 3 Biliary Stent and Delivery System

FDA Recall Z-1585-05 · Posted 2005-09-24

Recall StatusTerminated
Product CodeFGE
Date Initiated2005-07-21
Date Terminated2008-10-23
Product Quantity102

Reason for Recall

Angiomed Material Review Board reviewing the manufacturing process observed an increase in the internal scrap rate (failure mode: broken stent struts) of nitinol stents and determined the root cause to be a manufacturing change from polymer to aluminum fixtures.

Firm Action

Recall notification letter dated July 21, 2005, was sent to Hospital and Physician consignees.

Root Cause

Other

Distribution Pattern

Nationwide