MHI-TM2000 Linear Accelerator System Product Usage: MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.
FDA Recall Z-1585-2018 · Posted None
Advertisement
Recalling FirmHitachi Ltd., Medical System Operations Group
Recall StatusTerminated
Product CodeIYE
Date Initiated2017-12-08
Date Terminated2019-04-11
Product Quantity1
Reason for Recall
Due to a system controller software anomaly, the patient positioning deviation correction may not be applied and may result in the wrong part of the patient being irradiated.
Firm Action
Hitachi sent a Field Safety Notice letter dated November 28, 2017 to affected customers. The letter identified the affected product, problem and actions to be taken. The notification letter includes instructions for the user on how to verify patient position even with the software issue. The letter also explains that correction software is being developed with a tentative release date of, "End of January, 2018.". A customer service representative will contact affected customers about installing the software. For questions contact your our Customer Service Representative.
Root Cause
Device Design
Distribution Pattern
US Distribution in the state of NY.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.