Alto DR (model 614) and Alto 2 DR (model 624) Implantable cardioverter Defibrillator. Ventricular antichycardia pacing. Dual chamber arrhythmia detection. Antibradycardia pacing. ELA medical 98, rue Maurice Amoux, 92120 Montrouge- FRANCE. These ICDs are designed to recognize and treat slow or fast Ventricular Tachycardia or Ventricular Fibrillation by continuously monitoring atrial and ventricular activity to identify persistent ventricular arrhythmias and to deliver appropriate therapies.
FDA Recall Z-1587-05 · Posted 2005-09-24
Advertisement
Recalling Firmela Medical Llc
Recall StatusTerminated
Product CodeMRM
Date Initiated2005-07-19
Date Terminated2007-01-28
Product Quantity392 total devices (23-model 614 & 369-model 624)
Reason for Recall
Devices manufactured prior to August 2004 may experience premature battery depletion and/or extended charge times which may affect the device's ability to function.
Firm Action
A Dear Doctor letter dated 07/20/05 was sent describing the issue and providing recommendation that physicians follow-up with patients every 3 months. Patients who have frequent episodes of ventricular fibrillation or are pacemaker dependend may require prophylactic explantation or more frequent visits. A list of affected serial numbers was provided to each physician.
Root Cause
Other
Distribution Pattern
Nationwide. AZ, CA, FL, IL, KY, LA, MD, MI, MO, NC, NJ, NM, NY, OH, PA, SC, TN, TX, WA and WI.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement