MAKOplasty Hip Array Kit Laminate 150230
FDA Recall Z-1588-2016 · Posted None
Advertisement
Recalling FirmMako Surgical Corporation
Recall StatusTerminated
Product CodeOLO
Date Initiated2015-11-25
Date Terminated2016-08-25
Product Quantity10 units.
Reason for Recall
Stryker Orthpaedics became aware that there is the potential for the 210540 MAKOplasty Hip Array Kit Laminate 150230 to list two part numbers for the RIO Base Array, which is item #5 on the product table.
Firm Action
On 12/09/2015, Product Field Action Letters and Product Accountability forms were sent to Branch/Agencies via overnight courier with a return receipt; Hospital Risk Management were sent Notification letters and a Product Recall Acknowledgement form with return receipt on 12/9;/2015.
Root Cause
Component change control
Distribution Pattern
CO, KY, MN, WA and WV
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.