Analyzer and Atellica IM 1600 Analyzer system for in vitro diagnostic testing of clinical specimens.

FDA Recall Z-1588-2018 · Posted None

Recall StatusTerminated
Product CodeJJE
Date Initiated2018-02-26
Date Terminated2020-05-12

Reason for Recall

When more than one immunoassay analyzer is connected to a single solution system, reporting and conversion factors may not be applied to all analyzers

Firm Action

Recall initiated 02/23/2018 Firm released an update software (SW) version (V), 1.12.1, which corrects the behavior. A Mandatory Update (MU) campaign was launched to install the new SW V on all systems, whether or not they are impacted by this behavior. The MU campaign will be tracked to 100% completion globally.

Root Cause

Software design

Distribution Pattern

FL and France Spain Sweden United Kingdom Germany