Biomet INNERVUE DISPOSABLE SCOPE KIT 1.2MM WITH CANNULA SET, PRODUCT NUMBER 922141, STERILE
FDA Recall Z-1589-2011 · Posted 2011-03-07
Advertisement
Recalling FirmBiomet, Inc.
Recall StatusTerminated
Product CodeHRX
Date Initiated2011-02-10
Date Terminated2013-08-27
Product Quantity264
Reason for Recall
The cannula could not be connected to the scope and scopes disengage during use. The outer hub of Innervue scopes may become detached from the inner hub of the scopes during assembly of the scope to the Innervue hand piece. This condition may prevent the Innervue scope from being securely attached to the Innervue hand piece.
Firm Action
On 2/10/2011, Biomet issued an "URGENT MEDICAL DEVICE NOTICE" to their Direct Accounts instructing customers to identify and discontinue use of the recalled products and return them to Biomet. Response forms were provided.
Root Cause
Process design
Distribution Pattern
Distribution USA nationwide and to the Netherlands
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.