Imactis CT-Navigation System, stereotaxic accessory for Computer Tomography (CT), systems, Model Numbers E85101LA and E85101LP
FDA Recall Z-1590-2026 · Posted 2026-03-20
Advertisement
Recalling FirmGE Medical Systems, LLC
Recall StatusOpen, Classified
Product CodeJAK
Date Initiated2026-02-03
Product Quantity125 units
Reason for Recall
There is a potential issue that can lead to discordance between the simulated needle trajectory/tip position and the actual needle trajectory/tip position.
Firm Action
An URGENT MEDICAL DEVICE CORRECTION notification letter dated 2/3/26 was sent to customers.
Actions to be taken by Customer / User
You can continue to use your system if you DO NOT use needles with metal handles or metal hubs.
A User Manual addendum is included with this letter (see Appendix A) that includes specific instructions to avoid the use of needles with metal handles or metal hubs.
Please place a copy of the addendum in Appendix A with the User Manual for each Imactis CT Navigation" system in your facility.
Please ensure all potential users in your facility are made aware of this safety notification and the required actions.
Please retain this document for your records.
Please complete and return the attached acknowledgement form to [email protected] or use the QR code to submit your response electronically.
If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Root Cause
Device Design
Distribution Pattern
Worldwide distribution - US Nationwide.