Straumann coDiagnostiX, dental implant planning system. Software Version 7.0 including updates 7.0.2 and 7.0.2.10
FDA Recall Z-1592-2011 · Posted 2011-03-07
Advertisement
Recalling FirmStraumann Manufacturing Inc.
Recall StatusTerminated
Product CodeLLZ
Date Initiated2010-09-30
Date Terminated2011-09-19
Product Quantity20 systems
Reason for Recall
Software: Surgical Error may occur when the implant length is changed after positioning the surgical sleeve.
Firm Action
Straumann notified accounts of this correction by letter dated
September 30, 2010 and delivered via Federal Express Delivery Service. The upgrade is available on-line by logging into the coDiagnostiX system and enabling the automatic upgrade to occur.
Root Cause
Software design
Distribution Pattern
Nationwide
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.