ACUSON SC2000 Ultrasound imaging system with software version VB10C and using transesophageal (TEE) transducer Z6Ms, V5Ms or V7M. Model number: 10433816.
FDA Recall Z-1592-2016 · Posted None
Recalling FirmSiemens Medical Solutions USA, Inc.
Recall StatusTerminated
Product CodeIYN
Date Initiated2016-04-08
Date Terminated2017-01-18
Product Quantity87
Reason for Recall
While imaging with a transesophageal (TEE) transducer (Z6Ms, V5Ms, or V7M), user may lose the ability to control the color region of interest, the pulsed wave or continuous wave Doppler gate or cursor, the M-mode cursor, the 2D field of view, or the RES region of interest with the trackball.
Firm Action
Under Siemens update program US013/16/S a Customer Advisory Notification will be mailed to all affected customers by Certified mail with return receipt beginning March 30, 2016. Letters provide customers information on how to avoid the issue until the software patch is installed.
Root Cause
Software design
Distribution Pattern
Worldwide distribution. US nationwide, Austria, Canada, China, Germany, India, Italy, Spain, Sweden, and Switzerland.