Actim PROM dipstik, REF 30831ETUS; qualitative immunoassay rapid test for rupture of fetal membrane
FDA Recall Z-1592-2024 · Posted 2024-04-17
Advertisement
Recalling FirmACTIM OY
Recall StatusOpen, Classified
Product CodeOAM
Date Initiated2023-11-23
Product Quantity6471 kits
Reason for Recall
Interference testing has shown that the presence of personal lubricants in the vaginal specimen can interfere with the PROM test result by providing a false positive result.
Firm Action
Actim Oy issued a MEDICAL DEVICE RECALL notice to its sole distributor on 11/23/2023 via email. The notice explained the issue, potential risk, and requested that users refrain from using surgical lubricants when collecting the vaginal specimen and that vaginal specimens do not contain significant amounts of personal lubricant from patient's recent personal use.
The instructions for use will be updated according to the latest test results of surgical and personal lubricant interference. In the interim, a separate note will be included in the test kit regarding the correction of the interference claims and guidance of lubricant use.
Root Cause
Under Investigation by firm
Distribution Pattern
US Nationwide distribution in the state of Connecticut.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.