GORE VIBIL Biliary Endoprosthesis: stents, drains and dilators for the biliary ducts. Catalogue numbers: VH1008040, VH1010040, VN1010040, VN0808200 - Product Usage: intended for palliation of malignant strictures in the biliary tree.

FDA Recall Z-1593-2021 · Posted None

Recall StatusTerminated
Product CodeFGE
Date Initiated2021-04-05
Date Terminated2024-01-16
Product Quantity61 devices

Reason for Recall

The Gore post market surveillance team received a complaint that a GORE VIABIL Biliary Endoprosthesis appeared shorter than the labeled length.

Firm Action

On 04/08/2021, Gore issued an Urgent Medical Device Removal notice to customer via letter notifying them that a Gore Viabil Biliary Endoprosthesis 10 mm x 10 cm device appeared shorter than the labeled length.

Root Cause

Process change control

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, CT, FL, GA, HI, IL, KS, MD, MN, NE, NY, OK, RI, TX, VA and the countries of Portugal, Italy.