Open, Classified

Elecsys Troponin I STAT Immunoassay: Catalog# 05094798160 - Product Usage: intended to aid in the diagnosis of myocardial infarction.

FDA Recall Z-1595-2021 · Posted None

Recall StatusOpen, Classified
Product CodeMMI
Date Initiated2021-04-07
Product Quantity10, 489 kits

Reason for Recall

Recall reason is possible sporadic false negative conventional troponin results. Roche has confirmed an issue with running Elecsys Troponin I/Troponin I STAT and Elecsys proBNP II/proBNP II STAT on the cobas e 601 and 602 modules. When the proBNP II analyte is measured immediately before the Troponin I analyte on the same module, a decreased recovery on Troponin I was observed.

Firm Action

On 04/07/2021, Roche issued Urgent Medical Device Correction via fax notifying customers that Roche has confirmed an issue with running Elecsys Troponin I/Troponin I STAT and Elecsys proBNP II/ proBNP II STAT on the cobas e 601 and 602 modules. When the proBNP II analyte was measured immediately before the Troponin I analyte on the same module, a decreased recovery of Troponin I was observed.

Root Cause

Device Design

Distribution Pattern

US Nationwide distribution in the states of AL, AZ, CA, CT, FL, GA, ID, IL, IN,KS, KY, MD, ME, MO, NC, NE, NJ, NM, NY, OH, OK, SC, SD, TN, TX, VA, WA, WY.