Hospira Power cord use on Symbiq One-Channel infuser and Symbiq Two-Channel infuser. Power cord part number: HSP3306-E07. Used on List No. 16026 - Symbiq One Channel infuser; List No. 16027 - Symbiq Two-Channel infuser

FDA Recall Z-1596-2011 · Posted 2011-03-08

Recalling FirmHospira Inc
Recall StatusTerminated
Product CodeFRN
Date Initiated2010-10-27
Date Terminated2013-11-07
Product Quantity32,376 power cords

Reason for Recall

Complaints of broken, bent or missing prongs, charring, sparks, visible smoke, burnt smell have been reported on Symbiq AC Power cords.

Firm Action

Hospira Recall notification letters were sent by Federal Express on November 1, 2010.

Root Cause

Component design/selection

Distribution Pattern

Nationwide distribution and to Australia, Canada, and Malaysia.