Hospira Power cord use on Symbiq One-Channel infuser and Symbiq Two-Channel infuser. Power cord part number: HSP3306-E07. Used on List No. 16026 - Symbiq One Channel infuser; List No. 16027 - Symbiq Two-Channel infuser
FDA Recall Z-1596-2011 · Posted 2011-03-08
Advertisement
Recalling FirmHospira Inc
Recall StatusTerminated
Product CodeFRN
Date Initiated2010-10-27
Date Terminated2013-11-07
Product Quantity32,376 power cords
Reason for Recall
Complaints of broken, bent or missing prongs, charring, sparks, visible smoke, burnt smell have been reported on Symbiq AC Power cords.
Firm Action
Hospira Recall notification letters were sent by Federal Express on November 1, 2010.
Root Cause
Component design/selection
Distribution Pattern
Nationwide distribution and to Australia, Canada, and Malaysia.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.