StoneBreaker Single Use Probe, Catalog numbers: SBP-010500; SBP-010605; SBP-016500; SBP-016605; SBP-020425 GPN : G52607; G52605; G52606; G52609; G52608

FDA Recall Z-1596-2018 · Posted None

Recalling FirmCook Inc.
Recall StatusTerminated
Product CodeFFK
Date Initiated2018-01-31
Date Terminated2020-04-17
Product Quantity12,556

Reason for Recall

The interior of the product may not be sterilized to the appropriate Sterility Assurance Level when following the reprocessing instructions in the Instructions for Use.

Firm Action

On March 8, 2018 Cook Medical sent customers notification of an Urgent Medical Device Recall . Customers were instructed to complete the following actions: 1. Examine inventory immediately to determine if you have affected product, and quarantine affected product(s). Immediately cease all distribution and use of these products. 2.Return affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form. Used devices being returned should be double-bagged, placed inside an outer puncture free package with sufficient cushioning material to prevent movement between the secondary container and the outer package. An itemized list of the components must be placed in the container and the outer package marked Used Medical Device. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected product on hand, you must still complete the Acknowledgement and Receipt Form and return it via fax (812.339.7316) or email ([email protected]). 4. Immediately report adverse events to Cook Medical Customer Relations at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30 am and 5:00 pm (Eastern Time), or by email to: [email protected]. Customers with medical questions or concerns can contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235

Root Cause

Error in labeling

Distribution Pattern

US Nationwide, OUS countries: AE, AT, AU, BE, BG, BR, CA, CH, CL, CN, CO, CR, CZ, DE, DJ, DK, DO, EC, EE, EG, ES, FI, FJ, FR, GB, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IT, JO, KZ, LB, LT, LV, MA, MO, MX, MY, NI, NL, NO, NZ, PE, PH, PK, PL, PR, PT, QA, RO, RU, SA, SE, SG, SI, SK, SV, TH, TT, TW, UA, UY, XK, ZA