INFINITI Ultrasound Pack. Catalog Numbers: 8065751716, 8065751717, 8065751718, 8065751720, 8065751721, 8065751723. An ophthalmic microsurgical system indicated for phacoemulsification and removal of cataracts.

FDA Recall Z-1597-2016 · Posted 2016-05-03

Recalling FirmAlcon Research, Ltd.
Recall StatusTerminated
Product CodeHQC
Date Initiated2014-10-06
Date Terminated2017-03-31
Product Quantity6597 packs

Reason for Recall

The recalling firm received complaints related to leakage from the tubing near the white luer of the irrigation line for the INFINITI Ultrasound Packs.

Firm Action

The recalling firm notified affected customers on 10/6/2014 via a customer letter with an attached Medical Device Removal Response Form. On 10/23/2014, a second notification was sent to 76 customers who did not respond to the initial notification.

Root Cause

Process change control

Distribution Pattern

Nationwide Distribution.