Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

RayStation 4.5, 4.7, 4.9, 5, 6, 7, 8A, 8B, 9A, 9B, 10A, 10B, RayPlan 1, 2, 7, 8A, 8B, 9A, 10A, 10B including some of the Service Packs. Standalone software treatment planning system

FDA Recall Z-1597-2021 · Posted None

Recall StatusTerminated
Product CodeMUJ
Date Initiated2021-02-22
Date Terminated2023-05-01
Product Quantity116 units

Reason for Recall

For some LINAC types, merging clinical beams with beams of approximate dose may lead to the approximate dose erroneously labeled as clinical dose. Merge beams can be used manually, in scripting, or as part of the Automated breast planning feature. In some cases, when merging two beams where the first beam has clinical dose and the second beam has approximate dose, the dose of the resulting beam will be labeled as Clinical, although dose for some of the beam segments is still calculated with the SVD dose engine and it should be labeled Approximate: Mixed dose. The difference between approximate and final Clinical dose is in most cases small, but there can be body sites such as lung where the difference can be significant.

Firm Action

RaySearch issued Field Safety Notice, Medical Device Correction #74977 letter on 22 February 2021 via Email. Letter states reason for recall, health risk and action to take: ACTIONS TO BE TAKEN BY THE USER " Manual Merge beams in the 3D-CRT module Scenario: When merging a beam B1 with other beams (e.g., B2 and B3) in the 3D-CRT module. Action: If the target beam B1 has a clinical dose before the merge, make sure that all the beamsto be merged (B2 and B3) also have a clinical dose computed with the same dose engine. " Scripting when using the action MergeBeamSegments Scenario: When running a script containing the action MergeBeamSegments as below: beam_set.MergeBeamSegments(TargetBeamName="B1", MergeBeamNames=["B2", "B3"]) Action: If the target beam B1 has a clinical dose before running the script, make sure that the beams to be merged (B2 and B3) also have a clinical dose computed with the same dose engine. " Automated breast planning Scenario: When creating a treatment plan in the Automated breast planning module, the generated breast plan has two beams: A1 and A2. Actions: 1. Change module to Plan optimization. 2. Invalidate the dose for both the A1 and A2 beams by changing a beam property that can easily be restored (e.g., change the couch angle from 0 to 1). 3. Set the beam property back to the intended value (e.g., set the couch angle back to 0 for both beams). 4. Compute final dose. " Educate planning staff and all users about this workaround. " Inspect your product and identify all installed units with the above software version number(s). " Confirm that you have read and understood this notice by replying to the notification email. SOLUTION This issue will be resolved in the next version of RayStation/RayPlan, scheduled for market release in May 2021 (subject to market clearance in some markets). If customers wish to continue using versions of RayStation/RayPlan affected by this notice, all users must maintain awareness of this notice.

Root Cause

Software design

Distribution Pattern

US Nationwide distribution.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑