MAGNETOM Aera/Skyra Systems, whole body magnetic resonance imaging systems, Model numbers 10432914, 10432915. Docking station top cover (part no# 10591430) of the removable patient table.

FDA Recall Z-1598-2016 · Posted 2016-05-03

Recall StatusTerminated
Product CodeLNH
Date Initiated2016-04-13
Date Terminated2017-04-17
Product Quantity8 systems

Reason for Recall

The docking station top cover of the removable patient table may contain a magnetic metal plate. When docking/undocking the table, the magnetic top cover may become attracted to the magnet.

Firm Action

Siemens mailed to customers on April 13, 2016, a Customer Safety Advisory Notice, MR013/16/S, to inform them of the actions to take to prevent a possible hazard to persons and equipment. Customers were informed that they will be contacted by one of Siemens' service organization who will schedule an appointment to perform a correction, which is to exchange the top cover.

Root Cause

Vendor change control

Distribution Pattern

US nationwide distribution to OH, FL, TN, MN, OH, NY, VA, and MN.