6381 M-1 Cot Fastener, catalog no. 6381000000
FDA Recall Z-1599-2018 · Posted None
Advertisement
Recalling FirmStryker Medical Division of Stryker Corporation
Recall StatusTerminated
Product CodeFPO
Date Initiated2018-03-09
Date Terminated2019-04-09
Product Quantity3
Reason for Recall
The Operations Manual and Maintenance Manual are missing various warnings and contain incorrect information.
Firm Action
On March 9, 2018, the recalling firm notified affected distributors of the issue via their electronic product field action management system, who in turn notified their customers through written communication. Customers were informed of the issue and instructed to do the following:
1. Locate the operations manual shipped with each unit listed on the attached business reply form and replace it with the corrected manual included with the letter. Dispose of the incorrect manual.
2. Return the enclosed business reply form to confirm receipt of this notification to [email protected]
3. If you have loaned or sold any of the products listed in this letter, please forward a copy of this notice to the new users and advise the firm of their new location in the space provided on the business reply form.
4. If you have disposed of any of these units, advise the firm of their obsolescence by providing the firm with their serial number in the space provided on the business reply form.
Root Cause
Labeling Change Control
Distribution Pattern
Product distributed to the Netherlands
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.