Fisher Wallace Stimulator Models FW-100 and FW-200
FDA Recall Z-1600-2023 · Posted 2023-05-17
Advertisement
Recalling FirmFisher Wallace Laboratories Inc.
Recall StatusOpen, Classified
Product CodeQJQ
Date Initiated2023-04-06
Product Quantity43,489 total devices
Reason for Recall
If the batteries are installed incorrectly, the product may overheat, possibly leading to pain, discomfort, irritation, or even burns during use of the device.
Firm Action
An URGENT MEDICAL DEVICE CORRECTION dated 4/6/23 was sent to customers.
Short term actions to be taken by the User:
1. Always ensure the batteries are inserted correctly into the device (see above diagram).
2. Do not leave batteries in the device when the device is not in use (e.g. in between a morning and evening treatment).
3. Do not fall asleep while using the device because corrective actions cannot be taken when the user is asleep.
4. If the device exhibits excess heat, burning smell, or odor, turn off the device, and discontinue use. Contact Fisher Wallace Laboratories immediately.
The long term corrective action by the company is still being determined.
We request your action by replying to this email to confirm that you understand the information in this medical device correction and the subsequent actions needed to be taken.
Fisher Wallace apologizes for any inconvenience this may cause. If you have any questions about what to do with your product, please contact Fisher Wallace by telephone Monday - Friday, 10:00 AM to 6:00 PM, Eastern Time at (800) 692-4380 or by email at [email protected].
Root Cause
Nonconforming Material/Component
Distribution Pattern
US Nationwide. Global Distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.