Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Distal Centralizer, Exeter II Hip System. Catalog number - 0920-2-920. Intended to be used to centralize the femoral stem within the intramedullary canal. intended to be used with bone cement.

FDA Recall Z-1601-2010 · Posted 2010-05-13

Recall StatusTerminated
Product CodeJDI
Date Initiated2010-03-29
Date Terminated2012-09-10
Product Quantity24 units

Reason for Recall

Stryker became aware that device labels have incorrectly printed manufacture dates and thus incorrect expiration dates on the product label.

Firm Action

Urgent Product Recall letters and Product Accountability Forms have been sent via Fed Ex on March 29, 2010 with return receipt. Hospital Risk Management, Hospital Chief of Orthopaedics and Surgeon were also sent Notification letters and Product Recall Acknowledgment Form on March 29, 2010 with return receipt. The letters explained the issue, the potential hazards, and risk mitigation. The firm asked customers to examine their inventory and hospital locations to identify the affected products. All affected product should be retrieved and returned to their branch or agency warehouse for reconciliation. The attached Product Accountability form should be used to reconcile all product(s) and it should be faxed back to 201-831-6069.

Root Cause

Error in labeling

Distribution Pattern

Nationwide Distribution -- FL, MN, & OR.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑