Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

BD Vacutainer Partial Draw serum tubes (with or without gel).

FDA Recall Z-1602-05 · Posted 2004-10-28

Recall StatusTerminated
Product CodeJKA
Date Initiated2004-08-20
Product QuantityThere is no linkage to any particular lot of BD product

Reason for Recall

"Please be aware that this is not a new recall. This recall was initially classified in 2005 (Z-0052-05). The firm has taken action; but, due to administrative issues additional products are now being classified by the agency". There is an apparent bias in the determination of Total T3, Total T4, HBsAg, Folate BA, VB 12, BR, FSH and Cortisol hormones used to determine thyroid disease states, and a

Firm Action

The firm sent out a Technical Bulletin to all consignees of the specific instrument companies involved. Mailing initial were done 9/17/2004 and a revised mailing was sent 9/23/2004. Global communication will commense 10/5/2004.

Root Cause

Other

Distribution Pattern

Worldwide distribution.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑