Terminated

DERMALIFE Takara. Part #E30006-02 - Takara; Renew 2G, w/Pedestal. This is the Spa Jet product re-named for Takara Belmont, our New Jersey based distributor, identical to Part #E30006-01 other than the stickers on the unit.

FDA Recall Z-1605-2009 · Posted 2009-07-13

Recalling FirmSybaritic, Inc
Recall StatusTerminated
Product CodeIMB
Date Initiated2007-06-22
Date Terminated2012-07-05
Product Quantity10

Reason for Recall

A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA. Sybaritic, Inc. conducted a recall on DERMALIFE 2G equipment (SpaJet, Spa Oceana, Spa Feng Shui) because of IR Fan failure which resulted in the overheating and failure of the IR emitter and surrounding assembly.

Firm Action

Sybaritic, Inc. issued a "Safety Alert" dated April 15, 2008 informing "All Owners/Users of DermaLife 2G Equipment" of the affected device. The letter advised consignees on device use until the mandatory Filed Correction Kit is delivered. For further questions, contact Sybaritic, Inc. at 1-800-445-8418.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution -- US including states of CA, IA, MN, NE, NJ, NY, TX and WA and countries of Macedonia, Hungary, Greece, South Africa, Italy, Hong Kong, United Arab Emirates, Greece, Croatia, Australia, Kuwait, Turkey, Mexico, Belgium, England, Barbados, France, Japan, Lebanon, Moldova, Poland, Portugal, Romania, Russia, Spain and Thailand.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →