Terminated

Tapered HD Upgrade Set, REF TSKHDUS, Non-Sterile, Rx Only, Product Usage Usage: Surgical instruments used in preparation for dental implant placement

FDA Recall Z-1605-2015 · Posted 2015-05-12

Recall StatusTerminated
Product CodeDZE
Date Initiated2015-04-16
Date Terminated2015-08-07
Product Quantity48 units

Reason for Recall

BioHorizons Tapered HD Upgrade set was packaged with an incorrect drill component manufactured of stainless steel in place of the correct component manufactured from titanium.

Firm Action

BioHorizons sent an Urgent Medical Device Field Correction letter on 04/16/2015 to affected customers. Non-responding customers will be notified after 30 days. The letter identified the affected product, problem and actions to be taken. For questions contact Customer Care Department at 888-246-8338.

Root Cause

Process control

Distribution Pattern

US Nationwide in the states of: AL, FL, MS, VA, LA, CA, NC, TX, TN, OH, OK, VT, WA, MD, NY, SD, and SC.