Mariner RDX Polyaxial Head, REF: MX1-000020; Mariner RDX Extended Polyaxial Head, REF: MX1-000030; Mariner RDX Deformity Head, REF: MX1-000040; Mariner RDX Extended Deformity Head, REF: MX1-000050; Mariner RDX Trauma Head, REF: MX1-000060; Mariner RDX Extended Trauma Head. REF: MX1-000070, part of Mariner RDX Alpha Instruments and Implants, REF: AMRDX and Mariner RDX TD Alpha Implants, REF: AMRDXTD, which are pedicle screw systems.
FDA Recall Z-1605-2024 · Posted 2024-04-22
Advertisement
Recalling FirmSEASPINE ORTHOPEDICS CORPORATION
Recall StatusOpen, Classified
Product CodeNKB
Date Initiated2024-02-15
Product Quantity1579 Screw Heads
Reason for Recall
Screw heads, of pedicle screw systems, have thin lead-in thread that pose a risk of intraoperative cross-threading within the locking cap/screw head interface.
Firm Action
On 2/14/24 the firm phoned distributors and informed them of the recall issue and next steps. On 2/16/24 the firm sent a recall communication via email. Customers were asked to arrange for the return of these sets along with any additional Mariner RDX inventory using a firm provided return label.
Customers with questions or concerns can contact the Associate Product Manager, at [email protected]
Root Cause
Nonconforming Material/Component
Distribution Pattern
US Nationwide distribution in the states of MI, GA, VA, IN, FL.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.