Terminated

Dimension Total Prostate Specific Antigen (TPSA) Flex Reagent Cartridge (RF451). Siemens, Newark, DE 19714

FDA Recall Z-1606-2008 · Posted 2008-06-28

Recall StatusTerminated
Product CodeMTF
Date Initiated2008-03-24
Date Terminated2008-10-15
Product Quantity1631 cartons

Reason for Recall

False Results: The product may exhibit falsely elevated results. The affected lots demonstrate sporadic variability in recovery of calibrator and quality control; therefore this issue may not be readily detected upon calibration or during routine processing of Quality Control.

Firm Action

The recalling firm issued a letter (Urgent Field Safety Notice) dated 3/25/08 to inform their customers to discontinue use of any remaining inventory, retest using an alternate lot for those samples from post-prostatectomy patients which recovered positive TPSA results less than 1.0 ng/nL. The firm has requested that you call the them for no-charge replacement product at 1-800-441-9250 and/or for technical assistance.

Root Cause

Device Design

Distribution Pattern

The product was shipped to medical facilities nationwide. The product was also shipped to Germany, Singapore, Canada, Japan and AE