EMD Chemicals, Harleco Brand Ethanol Standard, 1.0 mg/ml; 1.0 mg Ethanol in 1 ml, 10 ampules; Item number 68991/95; in vitro diagnostic.
FDA Recall Z-1607-2008 · Posted 2008-09-16
Advertisement
Recalling FirmEMD Chemicals Inc
Recall StatusTerminated
Product CodeDNN
Date Initiated2008-03-27
Date Terminated2008-09-22
Product Quantity99 cases (10 ampules per case)
Reason for Recall
Incorrect concentration: 2 ampules measuring 2.0 mg/ml were found in a lot of 20 ampules of 1.0 mg/ml Ethanol Standard.
Firm Action
VWR International and Thermo Fisher Scientific were notified of the recall by e-mail on 3/27/08 and asked to provide end user lists so that recall letters could be sent to each customer. Customers were sent a letter commencing April 3, 2008, by first class mail.
Root Cause
Process control
Distribution Pattern
Nationwide Distribution.