DPM 6 Monitor in use with a Multi Parameter Module (MPM). Patient monitor used for monitoring, displaying, reviewing, storing, and alarming of multiple physiological parameters including ECG, arrhythmia detection, ST segment analysis, heart rate (HR), respiration rate (RESP), temperature, pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), carbon dioxide (CO2), anesthetic gas (AG), impedance cardiograph (ICG), bispectral index (BIS), cardiac output (C.O.) and respiration mechanics (RM)

FDA Recall Z-1611-2014 · Posted 2014-05-13

Recall StatusTerminated
Product CodeMHX
Date Initiated2014-03-05
Date Terminated2015-04-27
Product Quantity4 units

Reason for Recall

The MPM included a cardiac ST/Arrhythmia analysis feature which is not cleared for marketing.

Firm Action

Mindray DS Usa Inc. contacted their sole customer via phone on March 5, 2014. The affected products will receive software upgrade.

Root Cause

No Marketing Application

Distribution Pattern

US Nationwide Distribution in the state of OH