Open, Classified

Route 92 Medical, Full Length 070 Reperfusion System, REF: 7001, Rxonly, SterileEO. Used for Introduction of interventional devices into the neurovasculature.

FDA Recall Z-1611-2024 · Posted 2024-04-24

Recalling FirmROUTE 92 MEDICAL INC
Recall StatusOpen, Classified
Product CodeNRY
Date Initiated2024-03-08
Product Quantity750 units

Reason for Recall

Their is the potential for distal tip separation at the proximal marker band of the delivery catheters

Firm Action

On March 12, 2024 Route 92, issued a Urgent Medical Device recall notification to affected consignees via Mail and EMail. Route 92 asked consignees to take the following actions: Ensure all of your SUMMIT MAX inventory including the affected 070 Reperfusion Systems is in a secure, quarantined location. 2. Complete the Acknowledgement Form in Attachment 2 and return to Route 92 Medical by one of the following methods: [email protected]/Standard Mail - Attention Line: Tenzing 7 fiedl action, 155 Bovet Road Suite 100, San Mateo, CA 94402 Please return the Acknowledgement Form even if you no longer have the product subject to this recall. 4. Immediately return any affected products with the specific lot numbers listed in Attachment 1 to the address above using the original shipping container or request a new shipping container and pre-printed shipping label by email. 5. Keep this notice until all product subject to this recall has been returned. While processing your returns, please maintain a copy of this notice with the quarantined product subject to this recall and keep a copy for your records. 6. Separately, as part of SUMMIT MAX closure, you will be sending back your remaining SUMMIT MAX inventory if you haven t done so already.

Root Cause

Process control

Distribution Pattern

US: AL AZ CA DE FL GA IA IL IN KY LA MA MD MI MN MO NC ND NJ NY OH OR PA RI SD TN TX UT VA WV OUS: New Zealand

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →