Terumo Glidesheath (tm), 6 French, Nitinol wire, product code 15-1060. Used to facilitate placing a catheter through the skin into a vein or artery.
FDA Recall Z-1612-2013 · Posted 2013-06-27
Advertisement
Recalling FirmTerumo Medical Corporation
Recall StatusTerminated
Product CodeDYB
Date Initiated2013-05-30
Date Terminated2013-10-30
Product Quantity550 units
Reason for Recall
Medical devices were incorrectly labeled with extended expiration dates.
Firm Action
The recalling firm notified consignees by letter on 5/30/2013. The letter was flagged as an urgent product bulletin addressed to risk managers. The notification advised customers to immediately discontinue use and quarantine any available stock for subsequent instructions to be provided for pick-up and replacement.
Root Cause
Error in labeling
Distribution Pattern
Nationwide Distribution.