Terminated

Kinemax Stem Extenders, 80 mm, Product Usage: Kinemax Cemented Stem Extenders are modular components designed to be mated with tibial baseplates or femoral components of the Modular Rotating Hinge (MRH), Scorpio Total Stabilizer (TS) or Duracon TS knee replacement systems. The Kinemax Cemented Stem Extenders are available in three different lengths, 40 mm (Part #6476-8-250), 80mm (Part #6476-8-260) and 155mm (Part #6476-8-270). The 40 mm cemented stem extender is not indicated for use in the MRH system; however, it is indicated for use with the Scorpio TS and Duracon TS systems (LRKS/G Rev. 2). Both the 80mm and 155mm stem extenders are indicated for use in all three systems.

FDA Recall Z-1614-2015 · Posted 2015-05-14

Recall StatusTerminated
Product CodeJWH
Date Initiated2015-02-27
Date Terminated2016-10-13
Product Quantity4 units

Reason for Recall

Stryker received four customer complaints reporting that inside the box labeled part #6476-8-260 (Kinemax Stem Extenders, 80 mm) there was a different part #6476-8-250 (Kinemax Stem Extenders, 40 mm.

Firm Action

Stryker Orthopaedics an Urgent Medical Device Recall Letter/Notification Acknowledgement Form dated April 9, 2015 to their affected customers. The recall letter identified the affected product, problem and actions to be taken. Customers are asked to contact their Stryker Orthopaedics Sales Representative to arrange for return of the product if they have any of the affected items. Customers are asked to complete and return the acknowledgement form within 5 days by emailing to: [email protected] or faxed to: 1-855-251-3635. For question call 201-831-6365.

Root Cause

Mixed-up of materials/components

Distribution Pattern

US Nationwide distribution in the states of MA and RI.