Philips Respironics Trilogy Evo, Software Version 1.05.15.00. Continuous home-use ventilator device.

FDA Recall Z-1614-2026 · Posted 2026-03-26

Recall StatusOpen, Classified
Product CodeCBK
Date Initiated2026-03-02
Product Quantity113,717 units

Reason for Recall

Using non-pneumatic nebulizers with Trilogy Evo Platform ventilators may result in a discrepancy between the set tidal volume and the tidal volume received by the patient.

Firm Action

An UPDATED: "URGENT FIELD SAFETY NOTICE" dated 3/2/26 was mailed to consignees containing recall information for three different failure modes that can be experienced in all Trilogy Evo, Trilogy EVO O2, Trilogy Evo Universal, and Trilogy EV300 devices. The firm identified that using non-pneumatic nebulizers with Trilogy Evo Platform ventilators may result in a discrepancy between the set tidal volume and the tidal volume received by the patient. This can lead to under-delivery of therapy to the patient. Consignees are being instructed to update all devices with software version 1.05.15.00 through the My Philips for Professionals and In Center websites. The provided Appendix B contains the required instructions. Consignees are also instructed to review the latest version of the User Manual Addendum, communicate the safety notice to all in their organization who interact with affect devices, and forward the notice to those who received product that was further distributed. Consignees are to stop all use of non-pneumatic nebulizers, including vibrating mesh nebulizers, with all affected devices. Distributors should notify their customers of this recall with the provided notice and response form. DME and Homecare Customers can contact 1-800-345-6443 (option 2, option 1) for assistance. Hospital Customers can contact 1-800-722-9377 (option 2) for assistance.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide distribution: US (Nationwide) and OUS (Foreign) to countries of: Argentina, Austria, Australia, Bosnia and Herzegovina, Bangladesh, Belgium, Bulgaria, Bolivia, Brazil, Bahamas, Canada, Switzerland, Chile, China, Colombia, Czech Republic, Germany, Denmark, Algeria, Ecuador, Estonia, Egypt, Spain, Finland, France, Georgia, French Guiana, Gibraltar, Greece, Guatemala, Hong Kong, Croatia, Hungary, Indonesia, Ireland, Israel, India, Italy, Jersey, Jordan, Japan, Kenya, South Korea, Kuwait, Lebanon, Lithuania, Morocco, Montenegro, North Macedonia, Myanmar, Martinique, Malta, Mexico, Malaysia, Nigeria, Netherlands, Norway, Nepal, New Zealand, Oman, Panama, Peru, French Polynesia, Philippines, Pakistan, Poland, Portugal, Paraguay, Qatar, R¿union, Romania, Serbia, Russia, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Thailand, Tunisia, Turkey, Taiwan, Ukraine, United Arab Emirates, United Kingdom, Vietnam, Mayotte, South Africa, Zimbabwe.