ExploR 7x26mm Modular Radial Stem, Catalog number 11-210062. Elbow prosthesis orthopedic implant.
FDA Recall Z-1615-2014 · Posted 2014-05-13
Advertisement
Recalling FirmBiomet, Inc.
Recall StatusTerminated
Product CodeKWI
Date Initiated2014-04-22
Date Terminated2015-01-13
Product Quantity8 devices
Reason for Recall
This lot of the ExploR Modular Radial Stem may be missing the wedge ramp threads. If the wedge ramp threads are missing from the stem, the likelihood of the set screw loosening from the radial stem intraoperatively is increased. If the set screw loosens, a revision surgery may be necessary.
Firm Action
Biomet sent an URGENT MEDICAL DEVICE SAFETY ALERT dated April 22, 2014 to affected customers. The letter identified the affected product, problem and actions to be taken. Questions related to this notice should be directed to (574) 372-1570, Monday through Friday, 8 a.m. to 5 p.m.
Root Cause
Process control
Distribution Pattern
US Nationwide Distribution in the states of NY, MD, NC, SC, MI, OR, and NV.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.