GE Healthcare, Innova 3100 and Innova 4100 systems. The Digital Fluoroscopic Imaging System is indicated for use in diagnostic and interventional angiographic procedures of human anatomy. It is intended to replace image intensifier fluoroscopic systems in all diagnostic or interventional procedures.

FDA Recall Z-1616-2013 · Posted 2013-06-28

Recalling FirmGE Healthcare, LLC
Recall StatusTerminated
Product CodeOWB
Date Initiated2011-05-16
Date Terminated2014-03-13
Product Quantity48

Reason for Recall

GE Healthcare reported a potential safety issue due to temporary loss of X-Ray Imaging Modes during Collimator Auto-Calibration which affects Innova 3100 and Innova 4100 Systems.

Firm Action

A Customer Letter was not originally sent. GE Healthcare has retrospectively re-assessed the reportability of the field action and has determined that the action should be reported to the FDA. GE Healthcare has already updated all of the affected Innova 3100 and Innova 4100 systems that were distributed. As an effectiveness check, they have tracked all service requests for the field action to completion. These service requests have recorded the repair of the affected units. Exceptions to completion of the service request have been recorded as well. Records of field action completion have been retained per their processes and procedures. If you have any questions, please call 262-548-2757. For questions regarding this recall call 262-513-4122.

Root Cause

Software design

Distribution Pattern

Worldwide distribution - USA including AZ, CA, CO, FL, ID, IL, ME, NV,NY, NC, OK, TN, UT, and WI; and Internationally to Canada, Chile, France, Germany, Italy, Japan, Korea, Norway and Spain.