Terminated

Zimmer Kinectivo Technology Modular Neck, BB, 12/14 neck taper, sterile, Zimmer, Warsaw, IN; Catalog Number: 00-7848-022-01. Orthopedic implant used in total hip arthroplasty.

FDA Recall Z-1617-2009 · Posted 2009-07-27

Recalling FirmZimmer Inc.
Recall StatusTerminated
Product CodeLPH
Date Initiated2009-05-29
Date Terminated2009-12-22
Product Quantity147 of both products.

Reason for Recall

The corner tab of the inner compartment lid was not folded under before the outer lid was sealed, resulting in lack of assurance of sterility.

Firm Action

Zimmer, Inc. issued "Urgent: Device Recall" notices dated May 29, 2009 informing Consignees of the affected devices. Users were instructed to locate and return affected product and complete and return an Inventory Return Certification form via fax to the firm at 1-547-372-4265. For further questions, contact Zimmer, Inc. at 1-800-613-6131.

Root Cause

Packaging process control

Distribution Pattern

Worldwide Distribution -- United States, Brazil, Canada, Chile, Czech Republic, Finland, France, Germany, Italy, Korea, Lebanon, Switzerland and United Kingdom.