STA-Flex SUBTALAR SCREW, 12 mm x 18 mm, REF 982366, Sterile, Biomet Sports Medicine, 56 East Bell Drive, Warsaw, IN. Used to reduce excessive rearfoot pronation by blocking forward and downward displacement of the Talus, resulting in improved rearfoot alignment and foot function.
FDA Recall Z-1617-2011 · Posted 2011-03-09
Recalling FirmBiomet, Inc.
Recall StatusTerminated
Product CodeHWC
Date Initiated2011-01-21
Date Terminated2011-11-10
Product Quantity254 all products
Reason for Recall
The firm has initiated this recall following internal fatigue testing that shows possible initiation of fatigue cracking in implants under load.
Firm Action
The firm sent URGENT MEDICAL DEVICE RECALL NOTICES dated 1/21/2011 to distributors, implanting physicians, and medical facilities. The letter to the distributors and medical facilities states the recalled product should be immediately located and returned to Biomet. In addition, the "FAX Back Response Form" should be completed and returned to the number provided. If product has been further distributed, distributors are to contact hospital personnel via the enclosed "Dear Biomet Customer" notice. The letter to the implanting physicians informs the physician of the recall and lists the patients that were implanted with the device. The physician letter quotes from the surgical technique "the implant should also be removed if any postoperative complications occur". The letter to the implanting physicians does not recommend any action be taken with the devices that are implanted. The notice just makes the surgeons aware of the recall. Distributors and Physicians are to confirm receipt of the letter by contacting 800-348-9500, ext 3755 or 3756. Also, questions related to this matter should be directed to 574-371-3755 or 574-371-3756.
Root Cause
Pending
Distribution Pattern
Nationwide Distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.