20" Lead Wires, Replacement Part Number: 1067724-2
FDA Recall Z-1618-2024 · Posted 2024-04-23
Advertisement
Recalling FirmEBI, LLC
Recall StatusOpen, Classified
Product CodeLOF
Date Initiated2024-03-22
Product Quantity829 units (2 lead wires assembly)
Reason for Recall
Damage to the lead wire sheath (black cable) component and causing exposure of the wires, potentially lead to no therapeutic treatment signal being delivered to the electrodes, and result in delay of treatment
Firm Action
ZimVie issued Urgent Medical Device Correction letter to Patients, Physicians, Sales Representatives on 3/22/24 . Letter states reason for recall, health risk and action to take:
1. Review this notification for awareness of the contents.
2. Before each use, inspect your device s lead wires (black cable) for damage. If found to be
damaged, contact Customer Service at 1-800-526-2579 x 6000 for a complimentary
replacement.
3. Complete the Acknowledgement and Receipt Form either by scanning the QR code on the
postcard included with this letter and completing the online form or mailing back the completed
postcard. To complete your Acknowledgement, you will need to include your Unique Patient
Reference Number included on the top of the first page of this letter. This acknowledgement
should be completed even if you do not have affected product.
4. If you have further questions or concerns after reviewing this notice, please call customer service
at 1-800-526-2579 x 6000. Customer Service Representatives are available from 8:30 am to 5:00 pm (EST), Monday through Friday. At other times, please leave a clear message for a return call by the next business day. Alternatively, your questions may be emailed to [email protected].
Root Cause
Device Design
Distribution Pattern
US Nationwide distribution.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.