GENESYS Cross FT Suture Anchor with Two #2 (5 metric) Hi-FI Registered Trade Mark Sutures Rx Only STERILE EO CONMED LINVATEC REF CFBC-4502. Reattachment of soft tissue to bone in orthopedic surgical procedures

FDA Recall Z-1619-2013 · Posted 2013-07-01

Recall StatusTerminated
Product CodeMAI
Date Initiated2013-03-06
Date Terminated2013-07-22
Product Quantity60 each of both units

Reason for Recall

Incorrect blue/white suture was used.

Firm Action

ConMed Linvatec sent an "URGENT: Medical Device Recall Notification" letter by FedEx next day delivery on March 6, 2013, to notify all direct consignees of the recall. If the account no longer had the product in possess, they were asked to respond via a reply form or to contact ConMed Linvatec Customer Service to obtain a Service Request number for return and credit of unused unopened product. All returned product for this action would be identified, segregated and quarantined per the firm's procedures. Any returned product not part of the recall will be dispositioned in accordance with the firm's procedures. Please accept our sincere apology for any inconvenience this action may have caused . Should you have any questions please contact our Customer Service department at (800) 237-0169.

Root Cause

Mixed-up of materials/components

Distribution Pattern

US Distribution including the states of FL, KY, LA, MI, NC, and NY.