C-Reactive Protein Kit for use on the SPAplus, REF NK044.S.A, Rx only, UDI (01)05051700018456,
FDA Recall Z-1619-2020 · Posted 2020-03-31
Advertisement
Recalling FirmThe Binding Site Group, Ltd.
Recall StatusTerminated
Product CodeDCN
Date Initiated2019-04-25
Date Terminated2021-04-27
Product Quantity1 kit
Reason for Recall
A reagent in C reactive protein kits is not performing to specification and samples with known concentrations less-than-or-equal to 10 mg/L may be affected
Firm Action
On 4/25/2019, a Customer Notice was issued by the firm's UK office to firm's US office. The US office was instructed to provide this information to the end customer within 24 hours of receiving the notification along with a form to fill out and return.
The Customer Notice issued to the affected customer advises immediate quarantine and disposal of affected kits following local regulations. Customers were asked to pass on the customer notice and to all those who need to be aware within their organization or to any organization where potentially affected devices have been transferred. The customer notice requests the customer to complete, sign and return the e-back form. This confirms receipt of the Customer Notice, the number of kits disposed of and that the advice has been followed, along with a witness signature.
Firm requests that e-back form be returned to [email protected] or to a local firm representative within one week.
Should you require additional information, contact your local representative or Technical Support Group at UK: +44(0) 1214569696.
Root Cause
Nonconforming Material/Component
Distribution Pattern
U.S.: FL
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.