Terminated

FLAT DRAIN -10MM, CHARIKER-JETER DRESSING KITS (LARGE AND SMALL) -Product numbers 6680217 and 6680221 - Wound Management Product. Indicated for patients who wold benefit from a suction device particularly as the device may promote wound healing.

FDA Recall Z-1620-2008 · Posted 2008-08-06

Recall StatusTerminated
Product CodeBTA
Date Initiated2008-03-31
Date Terminated2008-11-18
Product Quantity4796

Reason for Recall

Incorrect labeling. Kit labels may incorrectly state that a flat drain is contained as a kit component while some kits may contain a round drain.

Firm Action

Consignees were notified of recall by letter (Correction and Removal Notification) sent to customers on 3/31/08. They were asked to check their stocks, identify, and return all affected units to Smith & Nephew according to the instructions. Consignees are to contact the Customer Care Center at 1-800-876-1261, if they have additional questions.

Root Cause

Labeling mix-ups

Distribution Pattern

Nationwide Distribution.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑