Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Hamilton G5 with software version between v2.41, v2.42, v2.50. Catalog number: 1590001 and 1590002. Anesthesiology: Hamilton G5 ventilator is designed for intensive care ventilation of adult and pediatric patients, and optionally infants and neonatal patients. The device is intended for use in the hospital and institutional environment where health care professionals provide patient care.

FDA Recall Z-1624-2016 · Posted 2016-05-09

Recall StatusTerminated
Product CodeCBK
Date Initiated2016-03-09
Date Terminated2017-06-29
Product Quantity1115

Reason for Recall

After performing the suctioning maneuver, including disconnecting the patient, suctioning , and reconnecting the patient, the preset pattern of ventilation many not continue as expected.

Firm Action

Hamilton Medical Inc. sent an" Medical Device Field Safety Corrective Action" letter dated March 7, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Hamilton Medical then had to compile a list of to whom they had distributed the affected ventilators and sent letters on April 6, 2016. Letters included the Medical Device Field Safety Corrective Action from Hamilton AG. Letters included immediate actions to be take by use and that reps will be visiting each site to upgrade ventilators to v 2.60 to alleviate the issue. Distributors will update software and will provide an updated Operator's manual as soon as it is available. Questions may be directed to 800-426-6331, Ext, 215.

Root Cause

Software design

Distribution Pattern

US distribution only.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑