-Scorpion Needle, For Use With Scorpion Suture Passer, (Blue). Ref. AR-13990N Lot 123263 -SureFire Scorpion Needle (purple). Ref. AR-13991N Lot 123263 and 122279 -Suture Passing Wire. Ref. AR-1255-18 Lot 123190 Manual surgical instrument for general use.
FDA Recall Z-1627-2013 · Posted 2013-07-01
Advertisement
Recalling FirmSurgical Instrument Service And Savings, Inc.
Recall StatusTerminated
Product CodeGDF
Date Initiated2013-05-03
Date Terminated2013-07-11
Product Quantity6 units
Reason for Recall
The seal which maintains a sterile barrier for reprocessed medical devices became compromised to the point where product may fall out of the pouch.
Firm Action
MEDISISS sent a Correction Immediate Action Request letter dated May 2, 2013, to all affected customes. The letter identified the product, the problem, and the action to be taken by the customer. The letter directed the customer to inspect the products prior to use for seal integrity as specified on the product labeling: "Sterile unless package has been damaged or opened. Do not use if package is damaged."
This will be followed up by a visit from the customer's sales representatives who are tasked with inspecting all the lots that may be remaining in the customer's inventory. The sales representative is directed to complete a "Verification of Seal Inspection" form and return it to MEDISISS Customer Relations. Devices with unacceptable seals are to be returned to MEDISISS. The sales representative is responsible for obtaining a shipping label from customer service to enable the devices to be returned to the recalling firm via Fed Ex.
Customers with questions were instructed to call 866-866-7477.
Root Cause
Packaging process control
Distribution Pattern
Nationwide Distribution.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.