Terminated

ABACUS SE (Single-Workstation Edition) and ABACUS ME (Multi-Workstation Edition) Product Usage: The ABACUS Software is a Windows-based order entry software application for comprehensive total parenteral nutrition (TPN) calculations and label printing.

FDA Recall Z-1628-2015 · Posted 2015-05-18

Recall StatusTerminated
Product CodeNEP
Date Initiated2015-04-08
Date Terminated2016-12-01
Product Quantity882

Reason for Recall

Baxter Corporation is conducting a field action for the ABACUS SE and ME models due to the possibility that the compounder will load the incorrect formula upon scanning the bag label when two or more different formula files have been created with the same order number.

Firm Action

Baxter sent a Safety Alert letter dated April 8, 2015 to affected customers via USPS first class mail. The letter identified the affected product, problem and actions to be taken. Customers were asked to 1) contact Baxter Technical Services if you need to restore the ABACUS databse, 2) foword a copy of this Safety Alert letter to other facilities or departments within your institution to ensure that those locations are awae of this action, and 3) complete the customer reply form and return it to Baxter by faxing to 224-270-5457 or scanning and e-mailing it to [email protected]. For questions contact the Center for One Baxter at 800-422-9837.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution - US Nationwide