iTotal PS Knee Replacement System, Model No. TPS-111-1111-010101, (ITOTAL PS IPOLY IMPLANT KIT LEFT) Product Usage: The iTotal CR Knee Replacement System is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis.
FDA Recall Z-1628-2020 · Posted None
Advertisement
Recalling FirmConformis, Inc.
Recall StatusTerminated
Product CodeJWH
Date Initiated2020-02-03
Date Terminated2020-12-02
Product Quantity2
Reason for Recall
A limited number of tibial trays were labeled with an incorrect serial number and were delivered to the wrong customer accounts.
Firm Action
Conformis Customer Service called three affected customers on February 4, 2020, to ask them to verify tibial tray serial numbers, and confirmed that the three kits in question had the incorrect tibial tray. These affected customers were asked to return the affected product to the firm.
Following this action, Conformis Sales Representatives were instructed to verify that all products within their control, including those already placed in hospitals, that were manufactured between September 1, 2019, and February 3, 2020, had tibial tray serial numbers that matched the patient's serial number. No additional affected product was found in the field.
Root Cause
Mixed-up of materials/components
Distribution Pattern
US Nationwide distribution in the states of LA, NY, TN.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.