Anspach Irrigation Clip 40 Product Usage: Irrigation clips provide a method of supplying controlled, cooling irrigation during cutting, shaping, and removal of bone, including bones of the skull and spine.
FDA Recall Z-1629-2014 · Posted 2014-05-16
Advertisement
Recalling FirmThe Anspach Effort, Inc.
Recall StatusTerminated
Product CodeFRN
Date Initiated2013-12-03
Date Terminated2016-04-18
Product Quantity150
Reason for Recall
The Irrigation Clips were assembled with the incorrect (smaller) clips.
Firm Action
The firm, Anspach, sent an URGENT: Medical Device Removal letter dated December 3, 2013 to its customer. The letter described the product, problem and actions to be taken. Should you have any queries, please do not hesitate to contact Aspach Customer Support at (800) 327-6887.
Root Cause
Mixed-up of materials/components
Distribution Pattern
US Nationwide Distribution in the state of Mississippi.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.