Terminated

JK Bariatric Beds, True Air Technologies, Inc.

FDA Recall Z-1630-2014 · Posted 2014-05-19

Recall StatusTerminated
Product CodeOSI
Date Initiated2014-03-18
Date Terminated2018-06-11
Product Quantity24 beds

Reason for Recall

The firm is recalling the products due to lack of design controls, lack of a Device Master Record, and failure to approve specific vendors.

Firm Action

SCM True Air Technologies LLC sent an Urgent Medical Device Recall letter to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to remove the affected product from use and return to True Air Technologies immediately. The units should be decontaminated and bagged with a certification of disinfection before return to True Air Technologies. True Air will provide replacement units to replace the returned units. Customers were instructed to refer to the Medical Device Recall Return Response Form attached for acknowledgement of receipt and contact information for this voluntary recall. Customers with questions should call 800-682-7163.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Nationwide Distribution including GA, KY, and MI.

Official Source

View FDA Recall →